Nedbrytningsprodukter av nervegasser analysert med væskekromatograf tilkoblet atmospheric pressure ionization elektrospray massespektrometer (HPLC-API-ES-MS) - del II: Negativ ion modus
dc.contributor | Hussain, Fatima | en_GB |
dc.date.accessioned | 2018-10-12T09:05:11Z | |
dc.date.available | 2018-10-12T09:05:11Z | |
dc.date.issued | 2002 | |
dc.identifier | 826 | |
dc.identifier.isbn | 82-464-0638-8 | en_GB |
dc.identifier.other | 2002/01729 | |
dc.identifier.uri | http://hdl.handle.net/20.500.12242/1535 | |
dc.description.abstract | Hydrolysis products of nerve agents have been analysed by HPLC coupled to an API-ES-MS in SIM mode. Parameters for the instrument were optimised for MPA, EPA, EMPA, IPMPA and PMPA. The instrument's detection limits for pure standards were 5 ng for the analytes, based on the condition that [M-H]- should be detected with a S/N > 3. The injection volume was 10 μl. | en_GB |
dc.language.iso | nob | en_GB |
dc.title | Nedbrytningsprodukter av nervegasser analysert med væskekromatograf tilkoblet atmospheric pressure ionization elektrospray massespektrometer (HPLC-API-ES-MS) - del II: Negativ ion modus | en_GB |
dc.subject.keyword | Nervegasser | en_GB |
dc.subject.keyword | Massespektroskopi | en_GB |
dc.source.issue | 2002/01729 | en_GB |
dc.source.pagenumber | 25 | en_GB |