dc.identifier.citationThon, Hansen, Austeng. Detection of Point Scatterers in Medical Ultrasound. IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control. 2021en_GB
dc.identifier.urihttp://hdl.handle.net/20.500.12242/2979
dc.descriptionThon, Stine Hverven; Hansen, Roy Edgar; Austeng, Andreas. Detection of Point Scatterers in Medical Ultrasound. IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control 2021en_GB
dc.description.abstractWe present an overview of the detection of point scatterers in ultrasound images and suggest strategies for evaluating and measuring the detection performance. We use synthetic aperture Field II simulations of a point scatterer in speckle background and evaluate how common imaging techniques affect point target detectability. We discuss how to compare different methods and calculate confidence intervals. In general, applying speckle reduction methods reduces the point detection performance. However, the results show that it is possible to smooth the speckle background and preserve relatively high performance with a suitable and optimized method. The different detection performances of the advanced beamforming methods Coherence Factor (CF), Phase Coherence Factor (PCF), and Capon’s Minimum Variance (MV) are presented and benchmarked with standard Delay-and-Sum (DAS). The results show that CF achieves slightly better detection performance than DAS for weak point scatterers, whereas PCF and MV perform worse than DAS. Choice of apodization window and adaptive aperture size affects the probability of detection. Results show that methods that preserve spatial resolution have better detection performance of point scatterers.en_GB
dc.language.isoenen_GB
dc.subjectUltralyden_GB
dc.subjectDeteksjonen_GB
dc.titleDetection of Point Scatterers in Medical Ultrasounden_GB
dc.date.updated2021-12-07T10:29:36Z
dc.identifier.cristinID1957182
dc.identifier.doi10.1109/TUFFC.2021.3129619
dc.source.issn0885-3010
dc.source.issn1525-8955
dc.type.documentJournal article
dc.relation.journalIEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control


Files in this item

This item appears in the following Collection(s)

Show simple item record